Carrier Gas in psi, Column Outlet and Vacuum in mmHg
A gas chromatograph asks for its inlet pressure in psi: the method file, the electronic pneumatic control screen and the regulator on the helium cylinder all speak that language. The other end of the column does not. Outlet conditions, detector specifications and every mass-spectrometer vacuum figure are published in millimetres of mercury, so working out what a method is really doing means putting both ends on the same scale.
What Head Pressure Is Actually Controlling
Linear velocity
Oven temperature
Outlet conditions
Gauge against absolute
Taking a Method Setting Through the Converter
Two directions come up during method development: turning a familiar inlet pressure into mercury units, and turning a published vacuum or outlet figure back into something the instrument screen understands.
Enter the head pressure from the method
Type the inlet setting — 8, 12.5, 20 — and the mercury figure appears as you go. If your regional keyboard produces a comma for the decimal mark, that is accepted just as readily as a dot.
Compare it against the outlet
Sea-level ambient is close to 760 mmHg, so a 12 psi head pressure (620.58 mmHg) means the column outlet is still at higher absolute pressure than the drop across the column itself.
Move the value into the notebook
The copy control beside each field puts the number on the clipboard on its own — unit stripped, spacing removed — so it lands tidily in a method record or a validation report. Ctrl + C with the cursor in a field does exactly the same.
Turn a vacuum spec back into psi
Press the swap arrows (↔) to work mmHg → psi, the direction you want when a detector datasheet quotes mercury. By hand it is a multiply by 0.019336775, and a source pressure of 1×10⁻⁵ mmHg comes out as 1.933677e-7 psi, shown in scientific notation because it is below one millionth.
Column, Inlet and Detector Conditions Side by Side
Indicative operating points across a capillary GC system, with the gauge pressures shown in both units and the absolute conditions at the far end of the column for context. Real values depend on column dimensions, carrier gas, oven programme and how the instrument was calibrated.
| Condition | Setup | Pressure (psi) | Pressure (mmHg) |
|---|---|---|---|
| Head pressure, 30 m × 0.32 mm | Helium, wide bore, low oven temperature | 4 psi | 206.86 mmHg |
| Head pressure, 30 m × 0.25 mm, hydrogen | Lower viscosity carrier at similar velocity | 5 psi | 258.57 mmHg |
| Head pressure, 30 m × 0.25 mm into an MS | Vacuum outlet, constant flow near 1 mL/min | 6 psi | 310.29 mmHg |
| Head pressure, 30 m × 0.25 mm, helium | FID outlet, oven at initial temperature | 8 psi | 413.72 mmHg |
| Same column at the end of the ramp | Constant flow held as viscosity rises | 15 psi | 775.72 mmHg |
| Head pressure, 60 m × 0.25 mm | Long column, same target velocity | 20 psi | 1 034.30 mmHg |
| Inlet leak-test pressurisation | Split vent closed, pressure held and watched | 25 psi | 1 292.87 mmHg |
| Carrier supply at the regulator | Cylinder second stage feeding the instrument | 60–80 psi | 3 102.90–4 137.19 mmHg |
| Column outlet at an FID (absolute) | Venting to the room at sea level | 14.696 psia | 760.00 mmHg |
That last pair is the anchor worth remembering: one standard atmosphere is 14.696 psi absolute and 760 mmHg, so any gauge pressure above about 14.7 psi has already exceeded a whole atmosphere. Against it, an ion source held near 1×10⁻⁵ mmHg is close to eight orders of magnitude lower — the reason a GC-MS transfer line, not the column, defines where the vacuum begins.
Built for Method-Development Arithmetic
Values that follow you while you retune
Both boxes accept typing and each updates the other immediately, so stepping a head pressure up in half-psi increments never means clearing a field.
Start from the mercury figure instead
The swap control reverses the pair to mmHg → psi, which is how a detector or vacuum specification usually arrives on the datasheet.
Torr and mbar in the same lists
Search either dropdown for the unit a pump or gauge manufacturer prefers; every pressure unit the app supports is offered on both sides of the pair.
Small numbers keep their digits
Results run to eight decimals and change to scientific notation below one millionth, which is exactly where ion-source and manifold pressures live.
Carrier Gas and Vacuum Questions
What head pressure gives roughly 30 cm/s of helium on a 30 m × 0.25 mm column?
On a cool oven venting to an FID it usually lands in single figures — around 8 psi, or 413.72 mmHg — but treat that as a starting point rather than a specification. Column length, internal diameter, film thickness, oven temperature and outlet conditions all move it, so confirm with an unretained marker or the instrument's own velocity calculation.
In constant flow mode, why does the inlet pressure climb during a run?
Because carrier-gas viscosity increases with temperature. To keep the same mass flow through the column as the oven ramps, electronic pneumatic control has to push harder — a method starting near 8 psi (413.72 mmHg) can finish close to 15 psi (775.72 mmHg). In constant pressure mode the pressure is held instead and the flow falls away as the oven heats, which is why retention times behave differently between the two modes.
Why does the same column need less head pressure on a GC-MS than on an FID?
The outlet is the difference. An FID exit sits at ambient, near 760 mmHg absolute, so the inlet must overcome that plus the column's own drop. A mass spectrometer holds the column exit at high vacuum, effectively zero on this scale, so a lower inlet pressure produces the same flow. Move a method between the two detectors without adjusting for vacuum outlet and every retention time shifts.
What delivery pressure should the carrier-gas regulator be set to?
Comfortably above the highest inlet pressure the method will ever call for, so the instrument's own control always has authority — commonly somewhere in the 60 to 80 psi region, which is 3 102.90 to 4 137.19 mmHg. Follow the instrument manual rather than a rule of thumb, and remember that supply pressure has no effect on chromatography beyond keeping the electronics in range: gas purity, traps and leak-free fittings decide the baseline.
How does a pressure-hold leak check on an inlet work?
The instrument seals the split and purge paths, pressurises the inlet to a set value — around 25 psi, or 1 292.87 mmHg, on many systems — then watches the decay for a minute or so. A tight inlet holds within the tolerance the manufacturer publishes, typically a fraction of a psi; a septum past its life or a loose column nut shows up as a steady fall. Half a psi lost is 25.86 mmHg, small on a gauge but plenty to spoil a baseline.
No comments yet. Be the first to comment!